| Model Number |
1005357H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the procedure was to perform electrophysiology in the coronary artery.The whisper ms guide wire was used in the patient however after the guide wire was removed, fluoroscopy confirmed that a portion of the guide wire remained in the patient.Several attempts were made to retrieve the separated portion with partial success.A small fragment migrated to the distal pulmonary artery and was determined to be too far to retrieve therefore it was left in the patient.Per the physician, the remaining portion was not expected to cause clinical issues for the patient.A delay in the procedure was reported.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to perform electrophysiology in the coronary artery.The whisper ms guide wire was used in the patient however after the guide wire was removed, fluoroscopy confirmed that a portion of the guide wire remained in the patient.Several attempts were made to retrieve the separated portion with partial success.A small fragment migrated to the distal pulmonary artery and was determined to be too far to retrieve therefore it was left in the patient.Per the physician, the remaining portion was not expected to cause clinical issues for the patient.A delay in the procedure was reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported difficulty could not be determined.The subsequent treatment appears to be related to circumstances of the procedure.Factors that may contribute to a material separation include, but are not limited to, tensile strength, corrosion, excessive force applied to device, interaction with accessory devices, and/or interaction with challenging anatomy.Based on the results of the complaint investigation, there is no indication a product quality issue with respect to manufacture, design or labeling of the device.
|
| |
|