| Model Number |
480440-06 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Shock (2072)
|
| Date of Event |
04/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that after a da vinci-assisted pancreatoduodenectomy procedure, the patient developed hemorrhagic shock secondary to postoperative bleeding, which required reoperation.The source and cause of the bleeding remain unclear.A synchroseal instrument was used during the procedure; however, the surgeon reported no intraoperative complications.The patient recovered and was discharged home.Additional information has been requested; however, no response has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no allegation that a malfunction of a da vinci system, instrument, or accessory occurred.A review of the energy log shows that the synchroseal instrument (lot number: k13240829-0118) was installed 4 times on universal surgical manipulator (usm) #3.The system logs show 16 coag events with no errors, 96 seal events with 1 ¿high initial starting impedance¿ error, 125 transect events with 3 ¿high initial starting impedance¿ errors; which can typically occur when the user tries to seal/transect excessive tissue.The system logs show a "recoverable error: instrument high initial starting impedance" at the same time as one of the 3 errors seen for the transect events above.
|
| |
|