| Model Number |
DISTAL DRILL GUIDE, Ø4.0mm |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, it was reported by a sales representative via (b)(6) that a 1268-100 targeting arm and 0315-100 drill guide are no longer targeting.Noticed during a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
-complaint allegation is confirmed.-visual evaluation of the 4.0mm distal drill guide, locking assembly, noted that the edge of the distal end of the shaft and green lever was chipped.Laser marking was observed to be faded.-functional testing was performed and noted that the spring did not function as required.The most likely cause of the reported failure can be attributed to wear and tear damage resulting from repeated insertion/removal/cleaning processes and extended use of the device in the field.
|
| |
|