• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Alcon Research, LLC - Huntington Clareon PanOptix Trifocal Hydrophobic Intraocular Lens; Lens, Multifocal Intraocular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Alcon Research, LLC - Huntington Clareon PanOptix Trifocal Hydrophobic Intraocular Lens; Lens, Multifocal Intraocular Back to Search Results
Model Number CNWTT0
Medical Device Problem Codes Break (1069); Failure to Unfold or Unwrap (1669); Material Twisted/Bent (2981)
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 08/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A nurse reported that during the presbyopic lens exchange surgery, the lens did not unfold well.No damage was observed on the lens during handling.The haptic was moderately folded in a different position, which was not uncommon.After implantation and unfolding of the iol, due to the final positioning, the surgeon noted that the anterior haptic to detach from its original location and the haptic adhered to the optic.Hence the lens was explanted during the same procedure and the corneal suture was performed.This caused the surgery to prolong and increased the endothelial stress.The corneal suture led to surgically induced astigmatism.The procedure was completed on the same day.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Clareon PanOptix Trifocal Hydrophobic Intraocular Lens
Common Device Name
Lens, Multifocal Intraocular
Manufacturer (Section D)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50473473
Report Number1119421-2026-01770
Device Sequence Number10112326
Product Code MFK
UDI-Device Identifier00380652408141
UDI-Public(01)00380652408141(17)290628(11)250630(10)16060414
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNWTT0
Device Catalogue NumberCNWTT0.210
Device Lot Number16060414
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/04/2026
Initial Report FDA Received Date08/27/2026
Date Device Manufactured06/30/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III Cartridge D.; PROVISC Ophthalmic ViscoSurgical Device.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-