• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medos International Sàrl INTRAFIX; Fastener, fixation, biodegradable, soft tissue

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medos International Sàrl INTRAFIX; Fastener, fixation, biodegradable, soft tissue Back to Search Results
Catalog Number 254802
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an anterior cruciate ligament reconstruction procedure, it was discovered that the sheath on the intrafx advbr scw9x23 w/lgshth device broke off.All the broken parts were removed.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: the product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found that the device is shown in used condition.Biological matter can be observed on the device.The sheath was found broken.The overall complaint was confirmed as the observed condition of the intrafx advbr scw9x23 w/lgshth would have contributed to the complained device issue.¿ based on the investigation findings, the potential cause is traced to the malleting action of the screw and levering the inserter can contribute to sheath breakage; as per instructions for use (ifu), the instructions for proper insertion are provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: there was no non-conformance regarding this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAFIX
Common Device Name
Fastener, fixation, biodegradable, soft tissue
Manufacturer (Section D)
Medos International Sàrl
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
Medos International SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
+1(908)386-3534
MDR Report Key50474503
Report Number1221934-2026-03836
Device Sequence Number10136109
Product Code MAI
UDI-Device Identifier10886705026746
UDI-Public(01)10886705026746(10)298L837(17)270430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254802
Device Lot Number298L837
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-