It was reported that during an anterior cruciate ligament reconstruction procedure, it was discovered that the sheath on the intrafx advbr scw9x23 w/lgshth device broke off.All the broken parts were removed.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: the product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found that the device is shown in used condition.Biological matter can be observed on the device.The sheath was found broken.The overall complaint was confirmed as the observed condition of the intrafx advbr scw9x23 w/lgshth would have contributed to the complained device issue.¿
based on the investigation findings, the potential cause is traced to the malleting action of the screw and levering the inserter can contribute to sheath breakage; as per instructions for use (ifu), the instructions for proper insertion are provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: there was no non-conformance regarding this lot.
|