• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Covidien Surgical Signia; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Covidien Surgical Signia; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGADAPTSTND
Medical Device Problem Codes Break (1069); Component or Accessory Incompatibility (2897); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, intraoperatively and prior to patient use, the adapter broke.A replacement adapter was obtained, and the issue was resolved without incident.Subsequently, an attempt was made to load a 60 mm reload; however, the reload could not be loaded, possibly due to damage to the adapter's center rod.Additionally, a radial reload was used for the first firing.Following firing, the radial reload was difficult to remove from the device.Although initial removal was challenging, the reload was eventually removed.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: sigradmt, tri 2.0 sigradmt rad med thk 12mm, (lot # n6d1572y) egia60amt, egia60amt egia 60 artic med thick sulu, (lot # p5l1030) sigphandle, sig power sigphandle handle, (serial # (b)(6)) sigpshell, sig power sigpshell control shell, (lot # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Signia
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Covidien Surgical
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
Covidien Surgical
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112--439
+1(800)646-4633
MDR Report Key50474764
Report Number1219930-2026-05461
Device Sequence Number18676025
Product Code GDW
UDI-Device Identifier10884521543805
UDI-Public(01)10884521543805(11)240521
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIGADAPTSTND
Device Catalogue NumberSIGADAPTSTND
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/21/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date08/27/2026
Date Device Manufactured05/21/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-