|
It was reported that during an unknown procedure that the clip applier was inserted into the patient and clips were deployed onto the target tissue as intended.After several clips had been deployed, the device jaws did not load the subsequent clip and the device was closed without a clip.Being mindful to fully depress and let the jaws open, we tried to load and fire a clip multiple times without success.There were still clips in the chamber, but they weren't loading in the jaws.Delay of case, but no patient harm.New clip applier was pulled to finish the case.There was 5mins of surgical delay was reported.
|
|
(b)(4).Date sent 8/27/2026.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained:
did device jam (not fire clips)? no.Did device not feed clips (clips not advance fully into jaws)? yes.Did device drop or eject clips? no.Was there any other issue noted with the clip firing (sideways feed clip, malformed clips, scissored clips, etc )? no.Did the clip not hold on the vessel or tissue once placed? yes.Was the device on a vessel/artery when it would not open? n/a.If yes, how was the device removed? (cut off, forced open) n/a.If the device was cut off, was there any damage to the vessel/tissue? n/a.
|