| Catalog Number |
PXSLIMLAN150T45 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the right peroneal artery using a lantern delivery microcatheter (lantern) and a glide catheter (4f).While removing the lantern from the patient, the physician observed that the lantern fractured at the midshaft within the glide catheter.The fracture was confirmed under fluoroscopy.Aspiration was applied to the glide catheter, and the glide catheter containing the fractured segment of the lantern was removed from the patient.The procedure was completed using a new lantern.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned lantern delivery microcatheter (lantern) confirmed that the catheter was fractured.If the lantern is manipulated at an angle or against resistance during use, damage such as a kink and subsequent fracture may occur.Based on the reported event, the root cause of this damage could not be determined.Penumbra devices are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|