Model Number 500004500500
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medbank mini system was powered off.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 14-jan-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system was powered off.A technical support specialist (tss) investigated the issue with the customer and determined that the beeping sound was coming from a surge protector near the power outlet that had automatically turned off.The tss advised reconnecting power to the medbank system and powering it on using the power button beneath the monitor.After the cabinet came back online, the drawers remained offline.The tss remotely accessed the system, closed and restarted the medbank application, verified the network adapter settings, and confirmed that the drawer offline warning was cleared and normal operation was restored.The system functioned as intended after the technical support specialist troubleshot the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedBank Mini
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50475923
Report Number2016493-2026-58171
Device Sequence Number18670049
Product Code BRY
UDI-Device Identifier10885403517181
UDI-Public(01)10885403517181
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500004500500
Device Catalogue Number1147-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1