A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 14-jan-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system was powered off.A technical support specialist (tss) investigated the issue with the customer and determined that the beeping sound was coming from a surge protector near the power outlet that had automatically turned off.The tss advised reconnecting power to the medbank system and powering it on using the power button beneath the monitor.After the cabinet came back online, the drawers remained offline.The tss remotely accessed the system, closed and restarted the medbank application, verified the network adapter settings, and confirmed that the drawer offline warning was cleared and normal operation was restored.The system functioned as intended after the technical support specialist troubleshot the issue.
|