AMO Puerto Rico Mfg. Inc. TECNIS IOL; Intraocular Lens Back to Search Results
Model Number ZCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during cataract surgery, a non-preloaded monofocal intraocular lens (iol) was implanted in the patient's eye.Following insertion of the iol, the optic was observed to be cracked.There was patient contact with the device.The affected iol was removed and replaced with an identical backup lens, and the procedure was completed successfully.The crack in the optic was first noted after lens insertion.The reporter indicated that the damage was not due to use error.It is unknown whether the tip of the cartridge was damaged.No patient injury, including capsule tear, was reported.No unplanned medical or surgical interventions, such as vitrectomy, incision enlargement, or sutures, were required.The current patient outcome is unknown.No further information was provided.
 
Additional Manufacturer Narrative
Section a2, a3a, a3b, a4, a5 and a6: unknown, information was requested but not provided.Section d6a: if implanted, give date: not applicable, as lens was removed/replaced during the same surgery.Section d6b: if explanted, give date: not applicable, as lens was removed/replaced during the same surgery.Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts were made to obtain additional information regarding the event; however, the information was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50476464
Report Number3012236936-2026-000253
Device Sequence Number10395985
Product Code HQL
UDI-Device Identifier05050474531697
UDI-Public(01)05050474531697(17)300428
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZCB00
Device Catalogue NumberZCB0000190
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/27/2026
Supplement Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Platinum cartridge, Lot: Unknown.
Patient SexUnknown
-
-