• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CVRx, Inc Barostim Neo2; Implantable Pulse Generator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CVRx, Inc Barostim Neo2; Implantable Pulse Generator Back to Search Results
Model Number 2104
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 07/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A barostim system was implanted on (b)(6) 2022, and a normal ipg replacement occurred on (b)(6) 2026.Following the procedure, the patient experienced a slight hematoma.Per the opinion of the physician, the root cause of the hematoma was the procedure for the battery change.On (b)(6) 2026, the hematoma was drained and it was noted that the patient was stable.As of (b)(6) 2026, it was confirmed that the hematoma resolved.
 
Additional Manufacturer Narrative
While analysis was unable to be performed as the device was not returned, per the opinion of the physician, the root cause of the hematoma was the procedure for the battery change.The device history and sterilization record for this device serial number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Cvrx id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Barostim Neo2
Common Device Name
Implantable Pulse Generator
Manufacturer (Section D)
CVRx, Inc
9201west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer (Section G)
CVRx, Inc
9201west broadway avenue
suite 650
minneapolis MN 55445
Manufacturer Contact
vincent mbibi
9201west broadway avenue
suite 650
minneapolis, MN 55445
MDR Report Key50476552
Report Number3007972010-2026-00100
Device Sequence Number10203397
Product Code DSR
UDI-Device Identifier00859144004623
UDI-Public(01)00859144004623(17)280112
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number2104
Device Catalogue Number100065-202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age67 YR
Patient SexMale
Patient Weight250 LB
-
-