|
Olympus Winter & Ibe GmbH Resection sheath, 8 mm, for 8.5 mm/26 Fr. outer sheath, ABS; Endoscope sheath, reusable
|
Back to Search Results |
|
| Model Number |
A42011A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the reusable endoscope sheath to the service center for a separate issue.During the device analysis, the service repair technician indicated that the ceramic tip (insulation beak) was broken.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for inspection.The root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: some kind of stress problem resulting in component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|