The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for the trilogy series.A field correction action(2021-05-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file¿ (b)(4) (trilogy risk matrix) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ allergy01, toxic01 reflect the safety risk assessment of design containing the affected pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.The reported failure of a battery failure is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review: device exceeds expected life.Dhr not required.
|