| Model Number |
1054260 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy 100 was returned to a third-party service center for evaluation.There was no harm or injury reported.There was no evidence the device was in patient use.During the evaluation a ventilator inoperative alarm condition indicating a motor failure was noted.The device logs were not available; therefore, the date of the most recent error codes occurrence could not be determined.At this time, the reported error code event date is based solely on the customer's information received and has not been confirmed through device evaluation.There was no documentation of any components replaced for this issue.There was no foam particles observed.No repairs were performed.The device was scrapped.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of a ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review: device exceeds expected life.Dhr not required.
|
| |
|
Search Alerts/Recalls
|