Model Number URF-V3R
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the endoeye flex deflectable videoscope for a separate issue.During the initial evaluation of the device it was observed that there was zero movement on the control knob.No patient involvement.
 
Additional Manufacturer Narrative
Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: stress and degradation.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Common Device Name
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50478473
Report Number9610595-2026-70210
Device Sequence Number15986590
Product Code FGB
UDI-Device Identifier04953170403392
UDI-Public04953170403392
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K181451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberURF-V3R
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/31/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-