| Model Number |
URF-V3R |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the endoeye flex deflectable videoscope for a separate issue.During the initial evaluation of the device it was observed that there was zero movement on the control knob.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: stress and degradation.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|