| Model Number |
SRP1/2R9E2 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a dental hygiene procedure, a student clinician was treating a class iv periodontal patient using a gracey 1/2 curette when the working tip fractured and separated from the instrument.The tip was immediately recovered.As a precautionary measure, an x-ray was taken to verify that no fragment remained embedded in the patient's subgingival tissue.The x-ray confirmed that no fragment was retained.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to hu-friedy for evaluation.Investigation determined that the failure is most likely associated with crack development and propagation due to mechanical fatigue and/or stress accumulated during use.In addition, observed cutting edge dulling and lack of timely resharpening may have contributed to increased force applied during clinical use, further promoting crack propagation and eventual fracture.The reprocessing of hu-friedy dental hand instruments and accessories states "inspect all instruments after the cleaning and rinsing step for corrosion, damaged surfaces, and impurities.Do not use damaged instruments."
the device history record for the subject lot was reviewed and no abnormalities were noted.
|
| |
|