Patient with history of large b-cell lymphoma was hospitalized on (b)(6) 2026 for confusion, worsening tremors, and hypervolemia.Patient was in need of mri and zoll respicardia was notified and interrogated the remedy system (b)(6) 2026.Testing revealed high impedance on 2 electrodes and providers were informed that the device is no longer considered mri conditional due to high impedance.During the course of hospitalization, cbc with differential was reported as abnormal on (b)(6) 2026 and blood cultures were obtained.Blood culture results on (b)(6) 2026 reported gram (+) cocci in pairs and chains identified as enterococcus gallinarum.Due to new finding of bacteremia and lead previously found to have high impedance and device is no longer considered mri conditional; provider made the decision to explant the remedy system.The system was explanted on (b)(6) 2026 and provider noted that inspection of the pocket showed no evidence of infection.
|