• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Synthes GmbH HANDLE BATTERY POWERED DRIVER; Instr,Surgical,Orthopedic,AC- Powered motor/access & attach

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Synthes GmbH HANDLE BATTERY POWERED DRIVER; Instr,Surgical,Orthopedic,AC- Powered motor/access & attach Back to Search Results
Catalog Number 05.000.008
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the front button not working.Problem found during routine testing.It was reported that there was no further information regarding the issue.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary.Service and repair evaluation: the customer reported that 05.000.008 front button not working.The repair technician reported that foreign substance protector/shield & motor, circuit board button failure, cable, electrical discolored & rust on motor.The cause of the issue is user error.The following parts were repaired: motor, protector/shield, circuit board & cable, electrical.The item will be repaired, put through final inspection, and will be returned to the customer upon completion of the service and repair process.Finalized service record will be archived in mds&r.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device history.Part # 05.000.008.Synthes lot # 007217.Supplier lot # 007217.Release to warehouse date: 05 jun 2018.Expiration date: na.Supplier: (b)(4).No ncr's generated during production.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDLE BATTERY POWERED DRIVER
Common Device Name
Instr,Surgical,Orthopedic,AC- Powered motor/access & attach
Manufacturer (Section D)
Synthes GmbH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
Monument
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50478732
Report Number8030965-2026-08593
Device Sequence Number10323605
Product Code HWE
UDI-Device Identifier10887587024585
UDI-Public(01)10887587024585(10)007217
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.000.008
Device Lot Number007217
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/22/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/04/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-