| Model Number |
CM1100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an incidental finding during product analysis confirming physical damage to the cardiomems i3 printed circuit board.Damaged area with a hole was discovered on a conductor on the i3 printed circuit board.Damage resulted in potential for exposed low voltage components.
|
| |
|
Additional Manufacturer Narrative
|
|
After performing multiple diagnostic tests and observing multiple failures, the unit was disassembled for further inspection.During disassembly, a damaged area with a hole was discovered on the pcb board.The damaged conductor was identified at u34.A standard power supply was connected to the unit.The unit was powered on with no sparking or thermal issues observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.
|
| |
|