Black Diamond Video, Inc.; BDV IDSS Custom Room Rack Back to Search Results
Model Number IDSS-SCC
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that the surgical displays to their bdv idss custom room rack were flashing and the wall monitors were not showing images.This event did not occur during a patient procedure.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the unit and found that the cabling needed to be resecured.To resolve the issue, the technician resecured the cabling, tested the unit, confirmed it to be operating according to specification, and returned to service.No additional issues have been reported.Udi related data clarification ¿ the lot/serial number is unknown, therefore, the applicable production identifiers were not included in the mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
BDV IDSS Custom Room Rack
Manufacturer (Section D)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer (Section G)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
+1(440)259-7401
MDR Report Key50480091
Report Number3008776287-2026-00597
Device Sequence Number10174033
Product Code KQM
UDI-Device Identifier00724995198473
UDI-Public(01)00724995198473
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIDSS-SCC
Device Catalogue NumberXQSILCC1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-