| Catalog Number |
04826876001 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable magnesium gen.2 results for 1 patient sample on a cobas 4000 c311 stand alone system.The initial result was 3.06 mg/dl with flag.The repeat result was 1.69 mg/dl.The repeat result was deemed correct.No questionable results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The alarm trace showed water reservoir too low, liquid level sensor in vacuum tank and liquid in vacuum tank alarms.It was found that the customer centrifuges samples at 5500 rpm for 5 minutes, which may be too forceful and too short of a time.The field service engineer (fse) found that the water tank dirty and cleaned and decontaminated it, and also found dried droplets on the surface around the reagent probe wash station.The fse performed a cell blank and a hardware check and cleaned and decontaminated the sample and reagent probes.The investigation is ongoing.
|
| |
|