| Model Number |
E1564 |
| Medical Device Problem Code |
Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleges "it was reported that during a leep procedure and bovie settings at 75, the device insulation melted, and a piece of it fell into the patient's cavity, which was fully retrieved.There was no patient injury.".
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.The device was used at 75 watts in the coag setting.Per the ifu, it states "always use the lowest power setting that achieves the desired surgical effect." and "maximum power for ball when used in coag setting is 30 watts.Exceeding maximum power settings may result in patient injury or product damage." as the device was used at settings much higher than the insgructions indicate, the root cause is user error.This should be considered the final report.If any additional relevant information is identified it will be submitted in a supplemental report.
|
| |
|