It was reported that the preloaded monofocal intraocular lens (iol) was implanted in the patient's right eye.During a postoperative examination, the patient experienced refractive surprise.The lens was explanted during a secondary surgical procedure and replaced with another dib00 intraocular lens of the same model with a diopter power of +18.5.No incision enlargement, vitrectomy, or sutures were required during the procedure.No additional information was provided.
|
|
Section a2, a3a, a3b, a4, a5 and a6: unknown, information was requested but not provided.Section b3: date of event: unknown, not provided, but the best estimate date is between (b)(6) 2026.Device evaluation:
visual inspection of the complaint device reveal that the lens was received cut in half and coated viscoelastic residue.No further issues were identified.The physical characteristics of the received lens was confirmed to match that of a dib00 model lens.No handpiece was received for evaluation.The complaint issues were not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to the manufacturing or design process.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed no other complaints were received for this po.Conclusion:
based on the complaint investigation results, no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|