• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Puerto Rico Mfg. Inc. TECNIS SIMPLICITY; Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Puerto Rico Mfg. Inc. TECNIS SIMPLICITY; Intraocular Lens Back to Search Results
Model Number DIB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the preloaded monofocal intraocular lens (iol) was implanted in the patient's right eye.During a postoperative examination, the patient experienced refractive surprise.The lens was explanted during a secondary surgical procedure and replaced with another dib00 intraocular lens of the same model with a diopter power of +18.5.No incision enlargement, vitrectomy, or sutures were required during the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Section a2, a3a, a3b, a4, a5 and a6: unknown, information was requested but not provided.Section b3: date of event: unknown, not provided, but the best estimate date is between (b)(6) 2026.Device evaluation: visual inspection of the complaint device reveal that the lens was received cut in half and coated viscoelastic residue.No further issues were identified.The physical characteristics of the received lens was confirmed to match that of a dib00 model lens.No handpiece was received for evaluation.The complaint issues were not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to the manufacturing or design process.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed no other complaints were received for this po.Conclusion: based on the complaint investigation results, no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50481141
Report Number3012236936-2026-000252
Device Sequence Number10396132
Product Code HQL
UDI-Device Identifier05050474731646
UDI-Public(01)05050474731646(17)270816
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDIB00
Device Catalogue NumberDIB00U0145
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/24/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-