Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary:
the device associated with this report was returned to depuy synthes for evaluation.The depuy synthes team forwarded the device to the service and repair site which conducted a visual and functional inspection of the returned device.The kincise 1 emphsys strght impct was visually and functionally assessed and determined to have failed the tool coupling assessment due to the locking collar fitting (stem deformed) damage, which prevented it from being fully inserted into the impactor, consistent with the adapter not being properly secured ¿ user error.No further action is required at this time since the reported issue is consistent with improper handling - user error.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the kincise 1 emphsys strght impct would have contributed to the complained device issue.¿
based on the investigation findings, it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history:
the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|