This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the product was not returned to depuy synthes, however a video was provided for evaluation.Manufacturing investigation results for vapr s90 w/ hand controls: based on the video supplied, we can confirm the video clearly demonstrates sparking occurring in the region of the manifold.This observation is consistent with the failure mode previously investigated.The observed sparking is also consistent with, and provides an explanation for the functional defect reported by the end user.The probable cause of the issue is a shorting event at the distal tip of the device, which can occur as a result of an incomplete adhesive seal.A design change has been identified as the corrective action required to eliminate recurrence of this failure mode.The overall complaint was confirmed as the observed condition of the vapr s90 w/ hand controls would contribute to the complained device issue.Based on the investigation findings, the potential root cause is traced to device design.This product issue was identified during the investigation by the supplier.This product issue will be addressed through the supplier quality system.Depuy synthes will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.Device history review: a manufacturing record evaluation was performed for the finished device u2602001 number, and no non-conformances were identified.
|