The device was returned to olympus for inspection, and the reported failure of "articulating the head of the scope, the screen is blacking out" was confirmed.The product analysis confirmed the following:
1.The image blacks out when bending section is angulated.2.The bending section has a bite/dent and unable to pass ring gauge.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the reported malfunction is traced to component failure: expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|