| Model Number |
IDSS-SCC |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that the touch panel used with the bdv idss custom room rack had a distorted image.No report of procedure involvement.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and confirmed the reported event.Following the technician¿s troubleshooting, he found that the capture card in the avc-x was defective.The technician replaced the capture card, tested the function and operation of the bdv idss custom room rack, confirmed it to be operating to specification, and returned the device to service.No additional issues have been reported.
|
| |
|