MAUDE Adverse Event Report: Medos International Sàrl FMS VUE II; Arthroscope

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medos International Sàrl FMS VUE II; Arthroscope Back to Search Results
Catalog Number 284004
Medical Device Problem Codes Increase in Pressure (1491); Improper Flow or Infusion (2954); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure, during the middle of the surgery, the fluid flow suddenly increased and began splashing from the incision site.Adjustments were made by decreasing and increasing the pressure; however, eventually only a small amount of fluid, at the level of dripping droplets, could be delivered even when pressure was applied.The procedure was completed successfully.There was no surgical delay and no patient harm reported.No further information is available.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative: the complaint device was not returned to depuy synthes, therefore unavailable for a physical evaluation.Given that no lot/serial number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot/serial number becomes available, the mre review will be performed.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.E1: the reporter¿s complete facility address was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE II
Common Device Name
Arthroscope
Manufacturer (Section D)
Medos International Sàrl
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
Medos International Sàrl
chemin-blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 
+1(908)386-3534
MDR Report Key50482891
Report Number1221934-2026-03849
Device Sequence Number10323748
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public(01)10886705027934
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171237
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number284004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/27/2026
Supplement Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-