Medline Industries, LP MEDLINE; MASK,OXYGEN,MEDIUM-CONCENTRATION,7,SC Back to Search Results
Catalog Number HUDRHO41U
Medical Device Problem Codes Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the 7 ft oxygen tubing fell off of the connector of the mask on 5 different masks today.The connection feels very loose and it is very easy for the oxygen tubing to fall away and thereby not deliver any oxygen to the patient.
 
Additional Manufacturer Narrative
The account reported that the 7 ft oxygen tubing fell off of the connector of the mask on 5 different masks, the connection feels very loose, and it is very easy for the oxygen tubing to fall away and thereby not deliver any oxygen to the patient.No sample or photo received for the investigation; a medline canada complaint history review for the past 12 months was performed and review of the previous three trend reports was completed for trending purposes, with trends identified and actions taken for detaching components.The reported issue was confirmed based on the reported lot number and its association with a recall regarding reported issues of tubing disconnecting from oxygen masks; the manufacturing site has halted production using the resin identified as the root cause of the recall.Our quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
MASK,OXYGEN,MEDIUM-CONCENTRATION,7,SC
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
brianna montgomery
three lakes drive
northfield, IL 60093
MDR Report Key50483127
Report Number1417592-2026-26434
Device Sequence Number16033169
Product Code BYG
UDI-Device Identifier10197344010646
UDI-Public10197344010646
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberHUDRHO41U
Device Lot Number25GOA760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2025
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-