| Catalog Number |
00055445010 |
| Medical Device Problem Code |
Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Following measurement of a sample associated with patient id (b)(6), a sample associated with patient id (b)(6) was subsequently measured.The user facility reported that when the barcode label for patient id (b)(6) was scanned, the system displayed the previously measured patient id (b)(6).The issue was reported to be associated with the gem 5000 system barcode reader/scanner.As a result, measurement data were uploaded to an online record associated with a different patient's sample slot.There was no adverse patient impact associated with this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
This is an initial report.The investigation is ongoing.A follow-up report will be submitted upon investigation completion.
|
| |
|