Product complaint # (b)(4)
date sent to the fda: 8/27/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained:
how many devices demonstrated the alleged deficiency? one defective device in this procedure.Did the event occur during one or multiple patient procedures? 1.What is the total number of procedures? 1.Have any of these events been previously reported to ethicon? if so, could you please provide the corresponding reference number(s)? only reported case, but doctor specified that this has multiple times with the exact same suture code in recent surgeries.During entry with (b)(6), needle snapped in half.If this event occurred during multiple procedures, would you kindly create a product complaint for each event and provide the corresponding reference number(s)?
did the needle fall into the patient? unknown.Were x-rays taken to locate the needle piece(s)? no.What measures were implemented to retrieve the broken piece? needle broke and was retrieved visually.Was there any additional tissue damage resulting from the search for the needle piece? no.Does a fragment of the needle remain in the patient¿s tissue? no.If yes, was more than half the needle retained in the patient? is there any plan in place to remove the needle piece in the future? if yes, could you please provide the scheduled date for the removal?
in which tissue structure was the broken needle located? unknown.What is the surgeon's opinion of the potential consequences for the patient? no consequence to the patient.Please provide the source or title of external person providing answers to follow-up:
(b)(6) (please refer to the attached email) materials manager.Please reply directly to this email.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|