| Model Number |
104232 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the system controller was showing a controller fault after the patient took a shower and got the controller wet.The patient was using the shower bag and the green running symbol was illuminated.Once the patient arrived at the emergency department (ed), the patient stated that the controller fault deteriorated to red alarms to no led screen.Green running symbol stayed illuminated.Controller exchanged performed in the ed.Changing the controller, batteries and battery clips resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
This event was previously reported under manufacturer report #2916596-2026-2918485.This product was added in addition to the product already reported.Section d4: device serial number and lot number were not provided, and expiration date and manufacture date (h4) cannot be determined.The primary udi number in d4 is reflective of all available information.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|