Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9733752, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: a medtronic representative went to the site to test the equipment.The side emitter was replaced and the system then functioned as intended.The side emitter, lot number: 603006782, was returned to the manufacturer for analysis.The complaint was verified.The returned side emitter faulted immediately when connected to the lat station.The emitter was then connected to the emtest and it was found to have a tcurrentfault on 4 of 12 coils.H6: codes b01, c13, and d02 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|