| Catalog Number |
MC1005T |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when inserting the second anchoring plate of the roi-c at c4/c5, it could not be advanced at all.When the plate was removed, it was found to be bent.It is suspected that the first plate had not been fully inserted, which may have caused interference with the second plate.The first plate was driven in again until it was fully seated.Another box was opened, and a replacement plate was used to continue the procedure.There were no reported patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|