| Model Number |
PRA-PROD-0007 |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6)2026, dose audit samples were irradiated for a routine dose audit, with sterility testing beginning (b)(6)2026 at (b)(6).By day 7, 5 of the 10 samples exhibited growth and the final 5 samples were inadvertently discarded before the full 14-day incubation period was completed which rendered the sterility test invalid.Samples from the same lote were retested and confirmed to pass sterility testing.Root cause of the failure has not yet been fully determined and the investigation is still on-going.No events have been reported related to patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Pramand is conducting an investigation on the invalid dose audit.Pramand consulted a sterilization expert who determined that the results of the test of sterility suggested a lab error and are considered invalid based upon available bioburden results from the test lab, test of sterility results, and identification of the positives.It was recommended that the test of sterility should be repeated due to the test lab's error in discarding the remaining samples from the dose audit.Pramand repeated the test of sterility with the same material from the same lot, and the lot met all specifications.The investigations of pramand, the sterilizer, and the test lab are still ongoing, with no root cause determined yet.
|
| |
|