| Catalog Number |
00515048200 |
| Medical Device Problem Codes |
Inability to Irrigate (1337); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery the device was not spraying and the battery was overheating.There was no patient impact or delay.During product evaluation it was found that the device leaked electrolyte.Due diligence is complete and there is no additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Functional evaluation of the returned product found the device did power on and function normally in both modes when connected to a lab battery pack.Visual evaluation found electrolyte had leaked inside of the pack, additionally, there did not appear to be damage to the wiring of the pack.No other damage was found.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause is attributed to a manufacturing issue exacerbated by a design issue.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1 telephone number: (b)(6).G2 country: china.
|
| |
|