| Model Number |
HexaVue IP |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported during a patient procedure the monitor to their hexavue ip custom room rack was experiencing ¿no video from guest port¿.The procedure was completed successfully and there was no report of procedure impact.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the monitor to the hexavue ip custom room rack subject of the reported event and found the universal port to the unit needed to be replaced and the cabling to the guest port needed to be retightened.While onsite, the technician proactively inspected the second monitor in the operating room and found that monitor also required repairs.To resolve the issue the technician replaced the monitors¿ universal ports and retightened the guest port cabling, tested the function and operation of the units, confirmed them to be operating to specification, and returned them to service.No additional issues have been reported.
|
| |
|