| Model Number |
RLM314K3 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported an intermittent ¿blinky¿ image on their vividimage 4k surgical display during a patient procedure.The procedure was completed successfully with a different display in the room.No report of injury or procedure delay.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the vividimage 4k surgical display and was unable to duplicate the reported event.The technician found that the fiber optic cable was pinched which likely caused the reported event.To resolve the issue the technician repositioned the fiber cable, tested the function and operation of the unit, confirmed it to be operating to specification, and returned the unit to service.The unit was installed in 2024 and is not under steris service agreement for preventive maintenance activities; the user facility is responsible for all maintenance activities.No additional issues have been reported.
|
| |
|