| Catalog Number |
NNU10LPT |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6), 2026, a patient underwent an ultrasound guided hydrothorax percutaneous drainage procedure using a navarre universal drainage catheter.During the procedure, the drainage catheter connector fractured.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
A patient underwent an ultrasound guided hydrothorax percutaneous drainage procedure using a navarre universal drainage catheter.Post the procedure on (b)(6) 2026, it was reported that the drainage catheter connector fractured.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f : the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: although the sample was not returned for evaluation, two electronic photos were provided and reviewed.The first photo shows a partial view of a drainage catheter attached to external connector.The catheter hub was noted to be cracked.The second photo shows the complete view of the catheter and fracture was not clearly visible on hub due to poor quality image.Therefore, the investigation is confirmed for the reported catheter hub fracture issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H.4.Device manufacture date: unknown.B5, g3, h6 (patient, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|