| Model Number |
IO-501 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Overheating of Device (1437); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's device was shutting off by itself and the battery was not lasting when they were outside of their home.The system hot error message was seen.As a result, the patient's oxygen levels decreased and they were taken to the hospital.No additional information is available at this time as multiple attempts to contact the patient were unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is estimated.The inogen team attempted to contact the patient for further information three times with no response.The unit came for system hot, and the complaint confirmed with data log and fan up and hitting accumulator probably due to dropped unit/high impact and that caused system heat.Reseat the fan fixed the issue.Column pressure is too high that caused by muffler and feed port assembly filled with dust and became blocked.
|
| |
|