Model Number IO-501
Medical Device Problem Codes Premature Discharge of Battery (1057); Overheating of Device (1437); Unexpected Shutdown (4019)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 08/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's device was shutting off by itself and the battery was not lasting when they were outside of their home.The system hot error message was seen.As a result, the patient's oxygen levels decreased and they were taken to the hospital.No additional information is available at this time as multiple attempts to contact the patient were unsuccessful.
 
Additional Manufacturer Narrative
B3: date of event is estimated.The inogen team attempted to contact the patient for further information three times with no response.The unit came for system hot, and the complaint confirmed with data log and fan up and hitting accumulator probably due to dropped unit/high impact and that caused system heat.Reseat the fan fixed the issue.Column pressure is too high that caused by muffler and feed port assembly filled with dust and became blocked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen Rove 6 Portable Oxygen Concentrator
Common Device Name
Inogen Rove 6 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50486321
Report Number3004672275-2026-00272
Device Sequence Number21355585
Product Code CAW
UDI-Device Identifier00817131020513
UDI-Public(01)00817131020513
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K230052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-501
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/14/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization;
Patient SexUnknown
-
-