| Model Number |
48360B-10 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The sureform 60 blue reload accessory was an incidental return with no allegation made against this product.No known impact or patient consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received a sureform (sf) 60 blue reload to perform failure analysis.The sf blue reload was found to have loose staples stuck in front of the knife.The reload was analyzed and found to have loose staples stuck in front of the knife.A visual inspection found the sf blue reload was partially fired approximately 90 percent through the firing process.There were 2 lodged staples, which could prevent the progression through the full firing trajectory.The probable root cause is attributed to loose staples from firing across staple lines or not having a full bundle of tissue between the jaws during use.
|
| |
|