| Model Number |
VICM5 13.2 (-5.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pupillary Block (2026); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault with elevated iop, pupil block., significant reduction of iridocorneal angles and shallowing of the ac and narrow angles.The lens exchanged for a shorter length lens and problem resolved.The cause of this event is the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Iop elevation from baseline , secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6-health effect clinical sign and symptoms code 4581:significant reduction of irido-corneal angles and shallowing of the ac.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|