| Model Number |
0998-00-0800-33 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6)centre & research.Event site postal code: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that during routine check the cardiosave intra-aortic balloon pump (iabp) was not switching on.No patient involved and no harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.Fse confirmed that the device did not power on and emitted beep sound.Found problem with display monitor to video generator board kit which have been replaced.All functional tests were performed and all calibrations were verified.Required calibration of vacuum and pressure regulators was done.Unit was tested with system trainer with balloon pumping and found working in satisfactory condition.
|
| |
|