| Model Number |
434893 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The patient had severe underlying cardiovascular disease, including multivessel coronary artery disease and reduced left ventricular function.During pci of the mid to distal lad, a type iii coronary perforation was reported after des placement.The perforation was treated with two pk papyrus covered stents and was reported to have been successfully sealed.During the same procedure, acute thrombosis/occlusion of both pk papyrus stents was reported, resulting in no-reflow and ventricular tachycardia/cardiac arrest.Following further intervention, final angiography reportedly showed timi 3 flow.The patient subsequently developed refractory shock and died later the same day.The cause of the acute stent thrombosis/occlusion and its potential contribution to the patient's death are under investigation.This report pertains to one of two pk papyrus devices used during the same procedure involving the same patient and clinical event.Since it could not be determined if only one or both pkps occluded, the second device is reported separately under manufacturer report number 8043892-2026-08114.
|
| |
|
Event or Problem Description
|
|
The patient had severe underlying cardiovascular disease, including multivessel coronary artery disease and reduced left ventricular function.During pci of the mid to distal lad, a type iii coronary perforation was reported after des placement.The perforation was treated with two pk papyrus covered stents and was reported to have been successfully sealed.During the same procedure, acute thrombosis/occlusion of both pk papyrus stents was reported, resulting in no-reflow and ventricular tachycardia/cardiac arrest.Following further intervention, final angiography reportedly showed timi 3 flow.The patient subsequently developed refractory shock and died later the same day.The cause of the acute stent thrombosis/occlusion and its potential contribution to the patient's death are under investigation.This report pertains to one of two pk papyrus devices used during the same procedure involving the same patient and clinical event.Since it could not be determined if only one or both pkps occluded, the second device is reported separately under manufacturer report number 8043892-2026-08114.
|
| |
|
Additional Manufacturer Narrative
|
|
New information: it was intended to treat a lesion with 99 % stenosis degree in the medial to distal lad.The affected pk papyrus stent system was not returned to the manufacturer and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report, discharge letter) and the production documentation was reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the devices were manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as a device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Search Alerts/Recalls
|