| Model Number |
PT-000409 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Battery Problem (2885)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the omnipod 5 controller/personal diabetes manager (pdm) charging port and cable melted after the cable was forcibly pulled on.The pdm would not hold charge following this and the patient's blood glucose level read as "high" overnight.No medical assistance was required.A replacement pdm and charging cable were issued to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned/received to date.If the device is received we will submit a supplemental with the investigation findings.We are unable to confirm the cause of the reported thermal event.The reported thermal device malfunction is associated with a personal diabetes manager manufactured after the correction for action res#91127 was implemented.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown
omnipod software app version: 3.1.6
operating system: n5004l-am-q-mv01602-06-01.06
hardware: n5004l
cgm sensor type: g7
please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|