Maquet Cardiovascular LLC Hemopro 3 Power Supply; Electrosurgical, Cutting & Coagulation & Accessories Back to Search Results
Model Number VH-6010
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Tw # (b)(4) updated sections: b4,b5,d9,g3,g6,h2,h3,h6,h11.The device was returned to the factory for evaluation on 08/10/2026.An investigation was conducted on 08/10/2026.A visual inspection was conducted.The power supply was returned outside its product packaging.An r&d engineering investigation was finalized on 09/25/2026.There was no visible damage observed.The led light was green.Voltage testing was conducted: no load voltage- less than 5.45v or over 5.55v- dmm measured 5.2 v.J300 3-pin molex connector was observed to be in place.J300 3-pin odu connector- the socket accepted a 1.18 mm pin gauge, whereas a known good connector rejected.Limit gauging (using meyers gage +/- 0.0025mm) revealed that two of the three connector sockets were structurally expanded and out of tolerance.Specifically, the compromised sockets accepted a 1.18 mm pin gauge, whereas a known good connector correctly rejected a pin of that size.Laboratory testing successfully replicated the failure mechanism.The insertion of an oversized test probe deformed the connector sockets, resulting in a subsequent loss of electrical contact with the standard-sized pins of the hp3 power cable.Based on the returned condition of the device as well as the evaluation results, the reported failure "failure to deliver energy" was confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Hemopro 3 Power Supply
Common Device Name
Electrosurgical, Cutting & Coagulation & Accessories
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50496627
Report Number2242352-2026-0000745
Device Sequence Number10631102
Product Code GEI
UDI-Device Identifier00607567701281
UDI-Public(01)00607567701281(11)251020
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K251238
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-6010
Device Catalogue NumberVH-6010
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/10/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Supplement Date Received by Manufacturer10/02/2026
Initial Report FDA Received Date08/28/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-