| Model Number |
HexaVue IP |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported loss of mouse and keyboard functionality during a patient procedure when switching between workstations through their hexavue ip custom room rack.The user facility subsequently reported loss of access to both systems, resulting in a procedure delay.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and determined that a loss of network connectivity had resulted in a loss of communication between the avc-x, usb-x, and mna components.The technician configured controls to the avc-x and the usb-x.The unit was tested, confirmed to be operating according to specification, and returned to service.No additional issues have been reported.
|
| |
|