Updated field: b4, g3, g6, h2, h6(investigation findings, investigation conclusions), h11
corrected fields: h6 (component codes).The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was not a maquet product.The extender and pressure tubing were also returned.A sensor output test was performed, and a pressure reading could not be obtained.The technician then used an optical light to detect any breaks in the sensors optical fiber and a break was observed within the y fitting.The optical fiber was found to be broken confirming the reported problems.We are unable to determine when this may have occurred.
|