• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0608
Medical Device Problem Codes Break (1069); Failure to Calibrate (2440); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated field: b4, g3, g6, h2, h6(investigation findings, investigation conclusions), h11 corrected fields: h6 (component codes).The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The returned sheath was not a maquet product.The extender and pressure tubing were also returned.A sensor output test was performed, and a pressure reading could not be obtained.The technician then used an optical light to detect any breaks in the sensors optical fiber and a break was observed within the y fitting.The optical fiber was found to be broken confirming the reported problems.We are unable to determine when this may have occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50497546
Report Number2248146-2026-0001470
Device Sequence Number10087883
Product Code DSP
UDI-Device Identifier10607567111728
UDI-Public(01)10607567111728(17)280915(10)3000500916(11)250915
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K122628
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0608
Device Catalogue Number0684-00-0608
Device Lot Number3000500916
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2026
Supplement Date Received by Manufacturer09/15/2026
Initial Report FDA Received Date08/28/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight136 LB
-
-