• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Perfusion Systems AUTOLOG IQ; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Perfusion Systems AUTOLOG IQ; APPARATUS, AUTOTRANSFUSION Back to Search Results
Model Number ATLGIQ
Medical Device Problem Codes Optical Problem (3001); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that during use of an autolog iq instrument, it was reported that it sent clear, translucent blood to the return bag. the instrument was replaced to complete the procedure.There was no adverse patient effect associated with this event.Medtronic received additional information that customer noted loud noises starting up.Medtronic received additional information that there were no visible or audible abnormalities with the unit.The issue with the unit was the optics fell out of specifications and this is a common issue.
 
Additional Manufacturer Narrative
Device evaluation summary: the reported issue the instrument sent clear, translucent blood to the return bag and made loud noises starting up was verified during service.The issue was resolved by calibrating optics.Post-repair testing was performed per specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOLOG IQ
Common Device Name
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
Perfusion Systems
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
Perfusion Systems
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50497866
Report Number2184009-2026-01324
Device Sequence Number13124258
Product Code CAC
UDI-Device Identifier00643169982864
UDI-Public(01)00643169982864(11)181128
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K181954
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATLGIQ
Device Catalogue NumberATLGIQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-