• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. ETHIBOND EXCEL; suture, nonabsorbable, synthetic, polyethylene

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. ETHIBOND EXCEL; suture, nonabsorbable, synthetic, polyethylene Back to Search Results
Catalog Number D6165
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 01/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by a spine surgeon that a patient underwent an unknown procedure on an unknown date and suture was used.Post-op, the patient came back with a wound dehiscence within two weeks post-op.The patient underwent surgical intervention to close the incision that dehisced.The surgeon stated they used this suture to perform an interrupted closure of fascia with a square knot technique, and claims that the knots are too slippery and don't seem to hold as well as a different manufactures suture that he was using previously.The surgeon believes this is why the knots fell apart post-op and caused a dehiscence.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information provided: i put that 5 eaches were affected, but the exact number is unknown.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.(b)(4) represents case 1 (b)(4) represents case 2 (b)(4) represents case 3 (b)(4) represents case 4 (b)(4) represents case 5 for each patient, please provide the following information: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.Date and name of index surgical procedure? the diagnosis and indication for the index surgical procedure? were any concomitant procedures performed? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? what was the tissue condition (normal, thin, calcified, fragile, diseased)? for the original intended closure, how were all layers closed? what tissue dehisced? onset date/time of dehiscence? (# post op days) how was the dehiscence managed? please describe any surgical intervention required for the wound dehiscence including date and findings.Please describe the appearance of the suture during the second procedure.Did the suture break? if so, where was the break noted (termination, middle, end)? did the suture pull out of the tissue? did the knots untie? were there any precipitating patient stress factors that led to the sutures untying, breakage or pulling out of the tissue? are photos available? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? surgeon¿s name? lot number? would the surgeon like to speak with ethicon medical safety and engineering via scheduled conference call regarding the events?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHIBOND EXCEL
Common Device Name
suture, nonabsorbable, synthetic, polyethylene
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50497881
Report Number2210968-2026-08753
Device Sequence Number9450918
Product Code GAT
UDI-Device Identifier10705031026117
UDI-Public10705031026117
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD6165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Supplement Date Received by Manufacturer09/14/2026
Initial Report FDA Received Date08/28/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-